Nanoduct™ Neonatal Sweat Analysis System Model 1030 Specifications

Readout 128 x 64 LCD graphic display (non-backlit). Supports up to 8 lines of 18 characters or numerals, with multi-lingual support (English, French, German, and Spanish).
Sound Alert and Alarm signals.
Keyboard ON, OFF, SELECT and ENTER keys.
Electrode Connection 6 pin locking medical connector to mate with induction/conductivity cell, cable.
Serial Outputs
RS-232 (ASCII format)
9-pin D-Sub male connector.
Data Protocol: 9600 Baud, 1 Start Bit, 8 Data Bits, No Parity, 1 Stop Bit.

Pin # Signal Description
1 N/C No Connection
2 RXD Receive Data (output)
3 TXD Transmit Data (input)
4 N/C No Connection
5 GND Signal Ground
6 N/C No Connection
7 N/C No Connection
8 CTS No Connection
9 /C No Connection

USB – Device
Type B Receptacle. 

Electrical Four AA Alkaline batteries (NEDA 15A, IEC LR6).
Typical solid-state, over-current circuit protection.
3.0 VDC lithium coin cell for the real time clock. Extended battery life of lithium cell is 75 years.
Sweat Induction Control Current profile controlled for use with Pilogel Iontophoretic disks with multiple fail-safe circuits to limit current. Nominal current is 0.5 (± 0.02) mA for 2.5 minutes (± 0.2 Sec.) Maximum fail-safe current limited to 5 mA.
Real Time Clock ± 2 minutes per year (battery backed).
Operating Temperature 15 to 30 °C (59 to 86 °F).
Storage Temperature

0 to 60 °C (32 to 140 °F).

Dimensions (H x W x D) 7.5 x 5 x 2 in (19.1 x 12.7 x 5.1 cm).
Weight

1.2 lb (0.5 kg).

Carrying Case (weight with instrument) 13.5 x 10.5 x 4 in (34.3 x 26.7 x 10.2 cm)
4.0 lb (1.8 kg).
Conductivity Readout mmol/L (equivalent NaCl).
Conductivity Range 3 to 200 mmol/L. *
Error 1% or less from 25 to 150 mmol/L (equivalent NaCl).
Initial Sweat Rate 0 to 50 g/m2/min.
Calibration Single point automatic calibration at 80 mmol/L (equivalent NaCl) using the AC-081 Calibration Plate.

* Solutions with a conductivity of less than 3 mmol/L are reported on the instrument as ≤ 3 mmol/L (equivalent NaCl).






Discovery Diagnostics